LogTag | USRIC-16 USB PDF Temperature Logger

23,80 ex. tax

  • –30 °C to +60 °C
  • 16,000 reading capacity
  • Built-in USB – no software or cradle needed
  • Auto-generated PDF report
  • Red/green visual alert indicators
USRIC-8 USB PDF Temperature Logger | LogTag | With Cap
USRIC-16 USB PDF Temperature Logger 23,80 ex. tax

for higher quantity pricing.

LogTag USRIC-16 – Single-Use USB PDF Temperature Logger

The LogTag USRIC-16 is a single-use USB temperature recorder ideal for monitoring sensitive shipments, clinical trials, and last-mile distribution. With no need for extra software or a cradle, it plugs directly into a USB port and automatically generates a detailed PDF report for immediate review.

Capable of storing up to 16,000 temperature readings across a range of –30 °C to +60 °C, it’s well suited for longer journeys. Visual indicators provide immediate feedback: green for OK, red for ALERT if any readings are outside set limits.

Designed for ease of use and wide compatibility, the USRIC-16 offers a dependable solution for one-way temperature monitoring and documentation.

Key features:

  • Temperature range: –30 °C to +60 °C

  • Stores up to 16,000 readings

  • Built-in USB – no cradle or software required

  • Automatically generates PDF report

  • Red/green status indicators

  • 6-month operating life after activation

  • 2-year shelf life (in hibernation mode)

Temperature Measurement Range -30°C to +60°C (-22°F to +140°F)
Storage Temperature Range 0°C to +40°C (32°F to +104°F)
Rated Temperature Reading Resolution Less than 0.1°C/F
LogTag Analyzer® currently displays to one decimal place of °C or °F. The native resolution is what is stored in the LogTag®
Sensor Reaction Time Typically less than 7 minutes (T90) in moving air (1m/s) EN12830:1999.
Recording Capacity 16,129 temperature readings. 56 days @ 5min logging, 168 days @ 15min logging.
Sampling Interval Logging interval is configurable from 1 minute to 18 hours. (Typically 5 minute logging).
Logging Start Options Push button start or specific date & time
Recording Indication Flashing ‘OK’ indicator / flashing ‘ALERT’ indicator
Download Time Typically in less than 20 seconds from time of insertion to availability of PDF report
Typically less than 10 seconds from time of insertion to availability of LTD file in LogTag® Analyzer (if configured)
Environmental IEC 60529: IP64 with USB cap fitted
Power Source CR2032 3V LiMnO2 Battery (Fixed)
Battery Life 2 years storage, followed by 6 months normal use
Real Time Clock Built-in real time clock
Rated accuracy ±25ppm @ 25°C (equivalent to 2.5 seconds/day)
Rated temperature coefficient is -0.034 ±0.006ppm/°C (I.e typically +/- 0.00294 seconds/day/°C)
Connection Interface USB 2.0, A-type plug
Software PDF Reader, LogTag® Analyzer 2.5 or higher.
Multi-language Support Multi-language support for English, Spanish, French, Russian and German
Size 93mm(H) x 54.5mm(W) x 8.6mm(T) including protective USB cap
Weight 31g
Case Material Polycarbonate